Unopette ProcedureThe Unopette disposable diluting pipette systemused to count WBCs is almost identical in shape andapplication to the Unopette system for RBC counts.The only major difference is that the reservoir containsa different diluent and the capillary pipette capacitydiffers (RBC 10 Fl and WBC 20 Fl). To assist you inperforming the Unopette procedure for WBCs, we willrefer to illustrations for the Unopette procedure forRBCs in this section.The Unopette procedure for counting white bloodcells is as follows:1. Puncture the diaphragm in the neck of thereservoir with the tip of the capillary pipetteshield. See figure 7-9.2. After you obtain free-flowing blood from alancet puncture of the finger, remove theprotective plastic shield from the capillarypipette. Hold the capillary pipette slightlyabove the horizontal and touch the tip to theblood source (fig. 7-10, view A). The pipettewill fill by capillary action. When blood reachesthe end of the capillary bore in the neck of thepipette, filling is complete and will stopautomatically. The amount of blood collectedby the capillary tube is 20 Fl. Wipe any bloodoff the outside of the capillary tube, making sureno blood is removed from inside the capillarypipette. (An alternative source of blood is athoroughly mixed fresh venous blood sampleobtained by venipuncture. See figure 7-10,view B.)3. With one hand, gently squeeze the reservoir toforce some air out, but do not expel any diluent(fig. 7-11). Maintain pressure on the reservoir.With the other hand, cover the upper opening ofthe capillary overflow chamber with your indexfinger and seat the capillary pipette holder in thereservoir neck (fig. 7-11).4. Release pressure on the reservoir and removeyour finger from the overflow chamber opening.Suction will draw the blood into the diluent inthe reservoir.5. Squeeze the reservoir gently two or three timesto rinse the capillary tube, forcing diluent intobut not out of the overflow chamber, releasingpressure each time to return diluent to thereservoir. Close the upper opening with yourindex finger and invert the unit several times tomix the blood sample and diluent. Seefigure 7-12.6. For specimen storage, cover the overflowchamber of the capillary tube with the capillaryshield.7. Immediately prior to cell counting, mix again bygentle inversion, taking care to cover the holewith your index finger.8. Place the coverglass on the hemacytometercounting chamber, making sure the coverglass isclean and grease-free. (Fingerprints must becompletely removed.)9. Remove the pipette from the reservoir. Squeezethe reservoir and reseat the pipette in the reverseposition. Release pressure to draw any fluid inthe capillary tube into the reservoir. Invert andfill the capillary pipette by gentle pressure on thereservoir. After discarding the first 3 drops, load(charge) the counting chamber of thehemacytometer by gently squeezing thereservoir while touching the tip of the pipetteagainst the edge of the coverglass and thesurface of the counting chamber (fig. 7-13). Aproperly loaded counting chamber should havea thin, even film of fluid under the coverglass(fig. 7-14, view A). Allow 3 minutes for thecells to settle. If fluid flows into the grooves(moats) at the edges of the chamber or if you seeair bubbles in the field, the chamber is floodedand must be cleaned with distilled water, driedwith lens tissue, and reloaded (fig. 7-14, viewB). If the chamber is underloaded, carefully addadditional fluid until properly loaded.10. Place the loaded hemacytometer into a petri dishwith a piece of dampened tissue to keep thehemacytometer from drying out (fig. 7-15).Allow 5 to 10 minutes for the cells to settle.11. Once the cells have settled, place thehemacytometer on the microscope. Using thehigh-power objective, count the WBCs in thefour corner fields of the hemacytometerchamber (fields A, B, C, and D of figure 7-16).Each field is composed of 16 small squares. Tocount the cells in each field, start in the upper leftsmall square and follow the pattern indicated bythe arrow in field B of figure 7-16. Count all ofthe cells within each square, including cellstouching the lines at the top and on the left. Donot count any cells that touch the lines on theright or at the bottom.7-18
Integrated Publishing, Inc. - A (SDVOSB) Service Disabled Veteran Owned Small Business